Accurate, repeatable and traceable temperature measurement for medical, research and industrial applications
Osensa phosphor-based fiber-optic probes provide high-resolution temperature measurement in environments where conventional electronic sensors may be unsafe, unreliable or affected by electromagnetic fields.
Fiber-optic temperature probes are used in MRI systems, thermal therapies, medical and veterinary research, RF and microwave heating, wireless charging evaluation, and industrial process monitoring. Calibration characterizes the complete measurement system so that the transmitter can convert the probe’s optical response into an accurate temperature reading.
Depending on the application, the measurement objective may be absolute temperature accuracy, relative temperature change, or both. Osensa systems can provide resolution as fine as 0.01°C and calibration accuracy of ±0.1°C, depending on the transmitter, probe, calibration range and operating conditions.
High-accuracy operation requires both a calibration table and an individual probe offset. These corrections serve different purposes and must match the probe and transmitter configuration.
The calibration table describes the optical response of a specific probe model used with a particular transmitter family over a defined temperature range. It contains temperature and decay-time data, together with the correction coefficients required by the transmitter. A table generally applies to probes of the same model, but Osensa can create a custom table for a specialized probe, range or application.
Normal manufacturing tolerances give each probe slightly different optical characteristics. The offset is a small positive or negative temperature correction that compensates for this probe-to-probe variation. It is recorded on the probe or calibration report and identifies the transmitter family for which it applies.
Both corrections are required for best accuracy. Using the wrong calibration table or omitting the probe offset can produce a significant measurement error; depending on the probe, the error may approach ±1°C.
When Osensa supplies a new system, the appropriate calibration table is normally loaded into the transmitter for the specified probe type. The probe offset is entered in OsensaView and uploaded to the assigned transmitter channel. During operation, the transmitter automatically applies the stored table and offset to calculate the corrected temperature.
If a probe is replaced, a probe model changes, or the probe is used with a different transmitter family, confirm that the correct calibration table and offset are available. If the probe serial number is known, contact info@osensa.com to request the applicable calibration information.
Osensa and accredited calibration laboratories use a precision metrology well together with a traceable platinum reference thermometer, precision readout, stable environmental conditions and specialized calibration software. The metrology well alone is not sufficient: the reference thermometer establishes the traceable temperature against which the Osensa system is compared.
Repeatability depends on placing the probe and reference thermometer in a uniform section of the well at a consistent insertion depth. Unused holes should be sealed where practical to reduce heat loss and outside-air circulation.
Settling time depends mainly on probe construction, tip mass, diameter, insertion depth and temperature step. Smaller probes generally stabilize more quickly. Record measurements only after both instruments are stable; inadequate settling time can introduce errors even when the calibration equipment itself is accurate.
Osensa can support:
When independent verification is required, Osensa can arrange testing through an ISO/IEC 17025-accredited calibration laboratory. Each probe is typically assigned to a specific transmitter channel and the complete system is verified against traceable standards. The laboratory then issues independent calibration documentation, which may support regulated applications or government submissions.
The table corrects the response of the probe model and transmitter family over the calibrated range. The offset corrects the remaining variation of the individual probe. Together, they provide the best system accuracy.
For applications requiring approximately ±0.1°C accuracy, annual recalibration is generally recommended. For applications requiring approximately ±1°C accuracy, a two-year interval may be acceptable. The final interval should follow the user's quality system, operating conditions, regulatory requirements and calibration history.
Yes. Osensa can develop tables for specialized probe designs, transmitter families, temperature ranges and application-specific requirements.
Check the probe label and calibration report first. If the information is unavailable, contact info@osensa.com with the probe serial number, probe model and transmitter model.